Clamp Meter
Calibration Adjustment
13
5. While shorting across the CAL keypad on the PCA, turn the rotary switch to the function to
adjust. See Figure 5.
6. When CAL shows on the display, remove the short.
This puts the Product into calibration mode.
7. When the real value shows on the display, apply the correct input signals from Ta b l e 5 to
the Product. For each function, the calibration step shown on the display will advance.
8. After each step, push on the front to confirm the calibration step, store the value, and go
to the next step.
9. Set the Calibrator to Standby after you complete adjustment of each function.
Note
If any calibration point is missing, Err is shown on the display.
After you push
, wait until the calibration step number advances before you change
the calibrator source. Some adjustment steps can take several seconds to execute
before the Product goes to the next step.
10. When calibration adjustment is complete, remove the 3.0 V supply.
The Product exits the calibration mode automatically.
11. Replace the batteries and battery door.
Figure 5. CAL Keyboard
Battery Contacts
CAL Keypad
calibration
seal