4
Measure a known voltage first to make sure that the
Product operates correctly.
Connect the common test lead before the live test lead
and remove the live test lead before the common test
lead.
Remove all probes, test leads, and accessories that are
not necessary for the measurement.
Before each use, examine the Product. Look for cracks or
missing pieces of the clamp housing or output cable
insulation. Also look for loose or weakened components.
Carefully examine the insulation around the jaws.
Do not use test leads if they are damaged. Examine the
test leads for damaged insulation, exposed metal, or if the
wear indicator shows. Check test lead continuity.
Disconnect power and discharge all high-voltage
capacitors before you measure resistance, continuity,
capacitance, or a diode junction.
Hold the Product behind the tactile barrier. See The
Product Figure 1 (
).
Do not use the Product if the wear indicator in the jaw
opening is not visible. See The Product Figure 2 (
).
Do not operate the Product with covers removed or the
case open. Hazardous voltage exposure is possible.

Other manuals for Fluke 319

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 319 and is the answer not in the manual?

Fluke 319 Specifications

General IconGeneral
CategoryClamp Meter
AC Voltage Range0.1 V to 600 V
DC Voltage Range0.1 V to 600 V
Continuity≤ 30 Ω
Safety RatingCAT III 600 V
Battery Type2 x AAA
AC Voltage Accuracy±1.5% + 5 digits
DisplayLCD
AC Current Accuracy±2% ±5 digits
DC Current Accuracy±2% ±5 digits