Electrical Specifications
The accuracy specification is defined as ±(% reading + digits) at
23 °C ± 5 °C.
Function Parameter 317 319
A
(A ac)
Range 40.00 A
600.0 A
40.00 A
600.0 A
1000 A
Resolution 0.01 A
0.1 A
0.01 A
0.1 A
1A
Accuracy 1.6 % ±6 digit
(50 Hz-60 Hz)
[40 A]
2.5 ±8 digit
(60 Hz-500 Hz)
[40 A]
1.5 % ±5 digit
(50 Hz-60 Hz)
[600 A]
2.5 ±5 digit
(60 Hz-500 Hz)
[600 A]
1.6 % ±6 digit
(50 Hz-60 Hz)
[40 A]
2.5 ±8 digit
(60 Hz-500 Hz)
[40 A]
1.5 % ±5 digit
(50 Hz-60 Hz)
[600/1000 A]
2.5 ±5 digit
(60 Hz-500 Hz)
[600/1000 A]
Crest factor
add 2 % into
spec for CF >2
3.0 Max @ 500 A
2.5 Max @ 600 A
3.0 Max @ 500 A
2.5 Max @ 600 A
1.4 Max @ 1000 A
AC Response rms rms