341
A
343A
b.
Set the 332B VOLTAGE RANGE to
1
00.
c.
Set the Model 343A
and
332B
voltage dials as
shown
in Figure
4-14
and
perform
the
indicated operations.
STEP 332B
VOLTAGE
DIALS
343A
VOLTAGE
DIALS
OPERATION
|
d.
10.00000
OOX.OOOO Adjust the 3328
dials for zero
(±10 uv) on the
845AR.
e.
10.00000
010.0000
Adjust DECK B
1
for
an 845AR
indication
within
10 uv of step
d.
1
20.00000 01 X.0000
Adjust the
332B
dials for zero
{
±10 uvi on the
845AR.
1
20.00000
020.0000 Adjust DECK B 2
for an 845AR
indication within
10
uv of step f.
h.
40.00000 03X.OOOO
Adjust the
332B
dials
for zero
(
±20 uv) on the
845AR.
i.
40.00000 040.0000
Adjust DECK B 4
for
an 845AR
indication within
20 uv of step h.
i>
60.00000
05X.0000 Adjust
the
332B
dials for zero
(
±20
uv)
on the
845AR.
k. 60.00000
060.0000
Adjust DECK
B 6
for
an 845
AR
indication within
20
uv of step
j.
1 .
80.00000
07X.OOOO Adjust
the
332B
dials for zero
(
±40 uv) on the
845AR.
m.
80.00000 080.0000
Adjust DECK B 8
for an 845AR
indication within
40 uv of step 1.
n.
100.00000 OOX.OOOO Adjust
the 332B
dials
for 2
ero
(
±40 uv) on the
845A R
o.
100.00000 0X0.0000
Adjust DECK B X
for an 845A R
indication within
40
uv of step
n.
Figure
4-14.
343A
LINEARITY
ADJUSTMENTS- DECK
B
a. Set the
Model 343A and 332B RANGE to
100.
b.
Set the Model 343A and 332B voltage dials as shown
in Figure
4-15
and perform the indicated operations.
STEP
332B VOLTAGE
DIALS
343A VOLTAGE
DIALS
OPERATION
c. 10.00000 OX.0000
Adjust
the 332B
dials for zero
(±10 uv
)
on the
845AR.
d.
10.00000 10.0000 Adjust DECK A 1
for
an 845A
R
indication within
10 uv of step c.
e. 20.00000
1 X.0000 Adjust
the 332B
dials for zero
(±10 uv) on the
845AR.
f. 20.00000 20.0000 Adjust DECK A 2
for an 845AR
indication
within
10 uv
of
step e.
9-
40.00000
3X.0000 Adjust the
332B
dials for
zero
(±20 uv) on the
845AR.
1
40.00000 40.0000 Adjust DECK
A
4
for
an 845A R
indication within
20 uv of step
g.
1
60.00000
5X.0000 Adjust the
332B
dials for zero
(±20 uv
)
on the
845AR.
j.
60.00000 60.0000 Adjust DECK A
6
for an 845A
R
indication
within
20 uv
of step
i.
1
80.00000 7X.0000
Adjust
the 332B
dials
for zero
(±40 uv) on the
845AR.
1
80.00000 80.0000 Adjust DECK A 8
for an 845AR
indication within
40 uv of step k.
m. 100.00000 9X.0000 Adjust the 332B
dials for zero
(±40 uv) on the
845AR
n.
100.00000 100.0000 Adjust DECK A10
for an 845AR
indication within
40 uv of step
m.
Figure
4-1
5. 343A LINEARITY ADJUSTMENTS-
DECK A