EasyManua.ls Logo

Fluke 341A - Model 343 A; Voltage Requirements

Fluke 341A
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
343A
STEP
VOLTAGE
DIAL
POSITION
OPERATION
|
c. OX.
0000
Record
voltage indication
within 10 uv.
d.
10.0000
Adjust
10V
LINEARITY
j
to
within 50
uv of step c.
e.
IX.0000
Record voltage
indication
within 40 uv.
f.
20.0000
Adjust 20V
LINEARITY
to within 70
uv of step
e.
9-
3X.0000
Record voltage
indication
within
40
uv.
h.
40.0000
Adjust
40V
LINEARITY
to
within 70
uv
of step
g.
i.
5X.OOOO
Record voltage
indication
within
40 uv.
j-
60.0000
Adjust 60V
LINEARITY
to within 70 uv of step i.
k.
7X.000Q
Record voltage
indication
within
40
uv.
1 .
80.0000
Adjust 80V
LINEARITY
to within
70 uv of step k.
m.
9X.0000
Record voltage
indication
within 40
uv.
n.
100-0000
Adjust 100V
LINEARITY
to within
70 uv of step
m.
Figure
4-10.
341 A
LINEARITY ADJUSTMENTS
b.
Set the
Model 341A
voltage dials as
shown
in
Figure
4-11. The output
voltages should be as
indicated
in
column 1 for
an
instrument that has just
been
cali-
brated. In
no case should
the output error
exceed
the
tolerances given in column
2,
which are
derived
from
instrument
specifications found in
Section I of
the
manual.
VOLTAGE
DIAL
POSITION
OUTPUT V
DC
1 2
10
0.00000 0
±0.00001
0
±0.00002
00.0000 0
±0.00002 0 ±0.0001
1000
000.000 0
±0.00006
0
±0.001
10
10.00000
10 ±0.0002 10
±0.001
10
5.00000 5
±0.0002 5 ±0.0005
100
100.0000
100
±0.002 100 ±0.01
100
50.0000
50 ±0.002 50
±0.005
100
10.0000 10
±0.0005 10 ±0.001
100
05.0000 5
±0.0003 5 ±0.0005
m
1000.000 1000 ±0.02 1000
±0.1
500.000
500 ±0.02 500
±0.05
1000
100.000 100
±0.005 100 ±0.01
100
050.000
50 ±0.003 50 ±0.005
c.
Adjust 10V
RANGE
ADJUST
for 10
±0.00003 volts
indication on the 895 A.
d.
Set
RANGE
switch to
100,
and
adjust 100V RANGE
ADJUST for
100
±0.0004 volts
indication on the
895A.
e. Set RANGE
switch to 1000 and
adjust 1000V
RANGE ADJUST for 1000
±0.004 volts indication
on the 895 A.
4-68. CALIBRATION
VERIFICATION
4-69. The following measurements should be
made
immediately following the range adjustments, paragraph
4-67.
Connect the
895A
to the OUTPUT
terminals, with
the
Model
341 A
negative OUTPUT
terminal con-
nected to chassis ground.
NOTE!
Use
the same 895A that was used
for
the range
adjustments.
Figure
4-11.
341 A
OUTPUT
VOLTAGE
REQUIREMENTS
c.
Calibration
of the Model 341 A is
complete. Set the
FUNCTION switch to
OFF and disconnect equip-
ment.
4-70.
MODEL 343A
PRELIMINARY CALIBRA-
TION
4-7
1 . Preliminary
calibration of the
Model
343A
consists
of
three main steps,
ordered as follows:
(1)
reference vol-
tage
adjustment,
(2)
preliminary zero
output and beta
string
adjustments, and
(3)
temperature
stabilization. Be-
fore
attempting calibration,
instrument performance
should
be examined
according to
the performance tests
of
para-
graph
4-24
to ensure
that no malfunctions
exist.
4-72. PRELIMINARY OPERATIONS
a.
Connect the Model 343A to the autotransformer,
and
adjust the
autotransformer for 115 volts
output.
b.
Set the Model 343A
controls as follows:
FUNCTION
RANGE
Voltage
dials
CURRENT LIMIT
ON
10
10.000000
Fully clockwise
4-15

Related product manuals