Effect of external
magnetic fields
of the magnitude of current in
adjacent wires)
Maximum conductor Size 40 mm
Maximum circuit voltage 300 V rms
Power Consumption 6 mW maximum
Automatic Power off Automatically turns off the power
approximately 10 minutes after
the last switch operation. The
beeper will turn on 15 seconds
before power off.
Power Supply CR2032 lithium battery
Battery Life Approximately 90 hours of
continuous operation
Dimensions 70 mm (W) x 176 mm (H) x 25
mm (D)
Weight Approximately 200 g
(including battery)
Safety Specifications
Category Rating
CAT III 300 V per IEC/EN61010-1,
and 61010-2-032 Pollution
Degree 2, indoor use
EMC
EN61326-1
Effect of Radiated Immunity
Rated accuracy +4.0 % in an electromagnetic field of 3 v/m.
Operating Altitude
2000 m
Accessories
C75
soft carrying case
Electrical Specifications
Reference Conditions: 23 ± 5 °C and 80 % RH maximum
Accuracy: ±(percent of reading + digit)
AC Current Measurement
Range Resolution Accuracy (50/60 Hz)
Maximum
Allowable
Current
3 mA 0.001 mA
30 mA 0.01 mA
1% + 5
30 A 0.01 A
60 A 0.1 A
0 to 50 A: 1 % + 5
50 to 60 A: 5 % + 5
60 A rms
Zero correction: Fractions smaller than approximately 0.01 mA
are calibrated to zero.