Performance Test Procedures
5
AC Current (iFlex Current with AC Voltage Simulation)
To verify ac current measurement with the iFlex, use the AC voltage simulation from the
calibrator:
1. Connect the normal output HI to the Calibrator Output HI to the iFlex 2 mm jack
flexible probe input terminal and the Calibrator Output LO goes to the black COM
jack . See Figure 1.
Figure 1. iFlex Current Probe Connections
2. Set the DUT to K (iFlex AC current measurement mode).
3. Set the calibrator to the voltage and frequency:
a. Verify that the DUT reads within the display limits.
b. Verify the DUT reading for the remaining steps for iFlex Current Probe (with
simulation).
0
/
+
5522A