Range Resolution
Frequency
(Hz)
Absolute / ±5 °C from calibration temperature Relative ±1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
±(ppm output + μV)
±(ppm output + mV)
220 V
[2]
100 μV
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
200 +4
75 + 1.5
45 + 0.6
70 + 1
120 + 2.5
700 + 16
4000 + 40
6000 + 80
220 + 4
80 + 1.5
47 + 0.6
75 + 1
130 + 2.5
800 + 16
4200 + 40
7000 + 80
230 + 4
85 + 1.5
50 + 0.6
77 + 1
140 + 2.5
850 + 16
4300 + 40
7500 + 80
240 + 4
90 + 1.5
52 + 0.6
80 + 1
150 + 2.5
900 + 16
4400 + 40
8000 + 80
200 + 4
75 + 1.5
35 + 0.6
60 + 1
110 + 2.5
500 + 16
3600 + 40
6500 + 80
220 + 4
80 + 1.5
40 + 0.6
65 + 1
120 + 2.5
600 + 16
3800 + 40
7000 + 80
1100 V
[1]
1 mV
15 - 50
50 - 1 k
240 + 16
55 + 3.5
260 + 16
60 + 3.5
280 + 16
65 + 3.5
300 + 16
70 + 3.5
240 + 16
40 + 3.5
260 + 16
45 + 3.5
5725A Amplifier:
1100 V
1 mV
40 - 1 k
1 k - 20 k
20 k - 30 k
75 + 4
105 + 6
230 + 11
80 + 4
125 + 6
360 + 11
85 + 4
135 + 6
440 + 11
90 + 4
165 + 6
600 + 11
50 + 4
85 + 6
160 + 11
55 + 4
105 + 6
320 + 11
750 V
30 k - 50 k
50 k - 100 k
230 + 11
600 + 45
360 + 11
1300 + 45
440 + 11
1600 + 45
600 + 11
2300 + 45
160 + 11
380 + 45
320 + 11
1200 + 45
Notes:
1. Maximum output 250 V from 15-50 Hz.
2. See Volt-Hertz capability in Figure A.
AC Voltage Secondary Performance Specifications and Operating Characteristics
Range
Frequency
(Hz)
Stability ±1 °C
[1]
24 Hours
Temperature Coefficient
Output Impedance
(Ω)
Maximum
Distortion
Bandwidth 10 Hz-
10 MHz
10 - 40 °C
0 - 10 °C and
40 - 50 °C
±μV ±μV / °C ±(% output + μV)
2.2 mV
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
5
5
2
2
3
3
5
5
0.05
0.05
0.05
0.1
0.2
0.3
0.4
0.5
0.05
0.05
0.05
0.1
0.2
0.3
0.4
0.5
50
0.05 + 10
0.035 + 10
0.035 + 10
0.035 + 10
0.035 + 30
0.3 + 30
0.3 + 30
2 + 50
22 mV
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
5
5
2
2
3
5
10
15
0.2
0.2
0.2
0.4
0.5
0.6
1
1
0.3
0.3
0.3
0.5
0.5
0.6
1
1
50
0.05 + 11
0.035 + 11
0.035 + 11
0.035 + 11
0.035 + 30
0.3 + 30
0.3 + 30
2 + 30
±(ppm output +
μ
V
) ±(ppm output
μ
V
) / °C
220 mV
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
80 + 15
12 + 2
10 + 2
10 + 2
20 + 4
100 + 10
200 + 20
2 + 1
2 + 1
2 + 1
15 + 2
15 + 4
80 + 5
80 + 5
80 + 5
2 + 1
2 + 1
2 + 1
15 + 2
15 + 4
80 + 5
80 + 5
80 + 5
50
0.05 + 16
0.035 + 16
0.035 + 16
0.035 + 16
0.035 + 30
0.3 + 30
0.3 + 30
1 + 30