EasyManuals Logo

Fluke 5730A Getting Started Manual

Fluke 5730A
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Multifunction Calibrator
Electrical Specifications
29
Range
Frequency
(Hz)
Stability ±1 °C
[1]
24 Hours
Temperature Coefficient
Output Impedance
(Ω)
Maximum
Distortion
Bandwidth 10 Hz-
10 MHz
10 - 40 °C
0 - 10 °C and
40 - 50 °C
±μV ±μV / °C ±(% output + μV)
Load Regulation
±(ppm output + μV)
2.2 V
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
80 + 15
12 + 4
15 + 5
15 + 5
30 + 10
70 + 20
150 + 50
50 + 10
15 + 5
2 + 1
10 + 2
10 + 4
80 + 15
80 + 40
80 + 100
50 + 10
15 + 5
5 + 2
15 + 4
20 + 4
80 + 15
80 + 40
80 + 100
10 + 2
10 + 2
10 + 4
30 + 10
120 + 16
300 ppm
600 ppm
1200 ppm
0.05 + 80
0.035 + 80
0.035 + 80
0.035 + 80
0.035 + 110
0.3 + 110
0.5 + 110
1 + 110
22 V
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
80 + 15
12 + 8
15 + 10
15 + 10
30 + 15
70 + 100
150 + 100
50 + 100
15 + 30
2 + 10
10 + 20
10 + 40
80 + 150
80 + 300
80 + 500
50 + 100
15 + 40
4 + 15
20 + 20
20 + 40
80 + 150
80 + 300
80 + 500
10 + 20
10 + 20
10 + 30
30 + 50
80 + 80
100 + 700
200 + 1100
600 + 3000
0.05 + 700
0.035 + 700
0.035 + 700
0.035 + 700
0.05 + 800
0.3 + 800
0.3 + 800
2 + 800
220 V
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 200
80 + 150
12 + 80
15 + 100
15 + 100
30 + 400
100 + 10,000
200 + 20,000
50 + 1000
15 + 300
2 + 80
10 + 100
10 + 500
80 + 600
80 + 800
80 + 1000
50 + 1000
15 + 300
4 + 80
20 + 100
20 + 500
80 + 600
80 + 800
80 + 1000
10 + 200
10 + 200
10 + 300
30 + .600
80 + 3,000
250 + 25,000
500 + 50,000
1000 + 110,000
0.05 + 10,000
0.05 + 10,000
0.05 + 10,000
0.05 + 10,000
0.2 + 50,000
1.5 + 50,000
1.5 + 50,000
3.5 + 100,000
±(ppm output + mV) ±(ppm output) / °C ±(ppm output + mV) ±(% output)
1100 V
15 - 50
50 - 1 k
150 + 0.5
20 + 0.5
50
2
50
5
10 + 2
10 + 1
0.15
0.07
1000
100
10
0
3.5 V
10000
50 Hz10 Hz 1 kHz 100 kHz 1 MHz 30 MHz
Voltage
Frequency
40 Hz 30 Hz
2.2 x 10
7
V-Hz
15 Hz
Volt-Hertz Capability
Figure A.

Table of Contents

Other manuals for Fluke 5730A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5730A and is the answer not in the manual?

Fluke 5730A Specifications

General IconGeneral
TypeMultifunction Calibrator
Resistance Range0 to 1 GΩ
AC Voltage Frequency10 Hz to 1 MHz
DisplayColor LCD
InterfacesRS-232, Ethernet
Weight18 kg
DC Voltage Range0 to ±1100 V
Dimensions172 mm H x 432 mm W x 457 mm D (6.8 in H x 17 in W x 18 in D)

Related product manuals