Remote Operation
SCPI Command Set 5
5-57
CH1 to CH6 Raw pulse streams from individual channels.
SUM456 Sum of channels 4 through 5.
MEAN456 Arithmetic mean of channels 4 through 6.
MEAN56 Arithmetic mean of channels 5 and 6.
EMUT On Every MUT channel. For example, on selecting a warm up
duration of 100 pulses, at least 100 pulses would have to be counted
on all selected MUTs before the warm-up sequence was considered
complete.
MAIN Equivalent pulse count of the power outputs (value calculated from
the “OUTPut:CONStant” meter constant value and active power
settings).
5-89. Test Sequence Tree
The test sequence commands specify the actions that will occur on sending “OUTPut ON”.
These parameters are applied in all operating modes except “FRUN”.
The source of pulses to count can be specified, as well as whether to count for a specific
period of time or counter value. The option to let the 6100 calculate the period (from energy)
is also available.
5-90. Test Duration
ENERgy:TEST:DURation(?) { SEConds| MINutes | HOURs | PPERiods | WH | KWH |
MWH }, <dnpd>
This command specifies the duration of the test sequence as either a time period, a counter
value, or a period defined in terms of energy.
SEConds As a time delay in seconds.
MINutes As a time delay in minutes.
HOURs As a time delay in hours.
PPERiods As pulse periods.
WH As accumulated energy.
KWH As accumulated energy.
MWH As accumulated energy.
The following are depreciated parameters from version I of the energy option:
COUNts As pulse counts.
ENERgy As accumulated energy.
Note: When energy is used (WH, KWH, MWH), the actual limits applied will be
determined by the meter constant selected for the pulse source.
5-91. Test Pulse Source
ENERgy:TEST:PSOUrce(?)
{ CH1|CH2|CH3|CH4|CH5|CH6|SUM456|MEAN456|MEAN56|EMUT|MAIN}
This command specifies the pulse source to use when determining the test completion
criteria.
The actual sources available will be determined by the active MUT source and reference
source selection.