6100B/6105A
Users Manual
4-12
output is on)
• Global settings Time/Date and GPIB settings cannot be changed in deferred mode
• Load/Save setup is not available in deferred mode
Note
Entry into Calibration mode automatically selects Direct mode.
4-25. Setting up Voltage and Current Waveforms
The following describes setting up voltage waveforms but applies equally to current.
Use to navigate to the Output Menu and use and until the voltage or current
channel to be set up is highlighted. Notice that the N-phase voltage channel is, by default,
limited to 33 V. The N-phase channel can be set to provide up to 1000 V if required. See
Figure 4-10.
XW Warning
To avoid electrical shock hazard, disconnect the 'N' phase
voltage Hi terminal from any 6140A Lo terminal before electing
to override the limit.
To override the limit:
1. Select the N-phase voltage channel in the Output Menu.
2. Select the Waveform Menu.
3. With the N-phase output set to OFF, check the Override Limit box.
gdw08.bmp
Figure 4-10. Channel Selection
Note
A channel must be enabled and
pressed for an output to appear at the
binding posts. If the output is already on but the active channel is not
enabled, pushing the Enable/Disable Channel softkey will cause the
output to appear at the relevant binding posts.