EasyManuals Logo

Fluke 62 MAX Calibration Information

Fluke 62 MAX
10 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #6 background imageLoading...
Page #6 background image
62 Max/62 Max +
Calibration Information
6
Performance Verification Tests
The subsequent performance tests verify the complete operation of the Product and check the accuracy of each
function against Product specifications. If the Product fails any part of the test, calibration adjustment and/or repair
is necessary. Product calibration can only be adjusted at the Fluke factory. To return your Product for calibration
adjustment, call one of the numbers in the “Contact Fluke” section of this document.
Read the entire Performance Tests section before you make any adjustment.
Required Equipment
Table 2 shows the required equipment necessary for the performance tests.
Note
Fluke recommends plate blackbodies as verification equipment because cavity blackbodies
have small field of view, which could increase uncertainty for a thermometer with low D:S. If
cavity blackbodies are used to verify the Product, make sure that the distance between the
cavity blackbody opening and the Product is equal to the diameter of the cavity blackbody
opening × 2. Where the temperature of the cavity tube near the outside is lower than the
temperature of the cavity, readings will be lower. Moving the 62Max closer to the cavity
opening may be necessary for the 62 Max to only measure the cavity temperature.
If other than Fluke blackbodies are used, the target should be checked for actual radiometric
output with a more accurate device, such as a Heitronic KT19-82.
Table 2. Required Equipment
Item Minimum Use Specifications Calibration Equipment
2.1 Blackbody
2.2 Blackbody
2.3 Blackbody
Range: -10 °C to 100 °C
Range: 100 °C to 500 °C
Range: 500 °C to 650 °C
Fluke 4180
Fluke 4181
Plate Blackbody with 10 cm to 15 cm
target.
[1]
[1] Calibrate with an IR thermometer that has low uncertainty such as KT19 and TRT.
Before the Performance Verification Tests
Before you do the performance verification tests:
Performance verification tests are to be done within the temperature of 23 ±3 °C and between 25 % and 60 %
relative humidity.
Make sure that all required calibration equipment has a valid calibration certificate and/or label.
Make sure that any item(s) substituted for the equipment in Table 2 meets or exceeds the stated minimum use
specifications.
The correct radiometric temperature will need to be determined before blackbodies are used to make
measurements.
Make sure that the required calibration equipment is allowed a warm-up period as per manufacturer
recommendations. See the documentation for the equipment.

Other manuals for Fluke 62 MAX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 62 MAX and is the answer not in the manual?

Fluke 62 MAX Specifications

General IconGeneral
TypeInfrared environment thermometer
PurposeOutdoor
PlacementHand
Display typeDigital
Product colorYellow
Display form factorRectangular
International Protection (IP) codeIP54
Accuracy1 °C
Response time500 ms
Spectral response8 - 14 µm
Min/Max temperature memoryYes
Temperature measurement range-30 - 500 °C
Temperature measurement unitsF, °C
CertificationEN 61326-1:2006 EN 61326-2:2006 EN/IEC 61010-1: 2001 EN 60825-1 Class II
Battery typeAA
Power sourceBattery
Number of batteries supported1
Operating altitude0 - 2000 m
Storage temperature (T-T)-20 - 60 °C
Operating temperature (T-T)0 - 50 °C
Operating relative humidity (H-H)10 - 90 %
Battery life10 h
Weight and Dimensions IconWeight and Dimensions
Depth75 mm
Width85 mm
Height175 mm
Weight255 g

Related product manuals