Calibration Information
2
Accuracy Specifications
Accuracy is specified for 1-year after calibration at
operating temperatures of 18 °C to 28 °C and is given
as:
± ( [ % of reading ] + [ counts ] )
MEASURE V dc:
Range: + 28 V max
Resolution: 1 mV
Input Impedance: 1 MΩ
Accuracy: ± (0.015 % of reading + 2 counts)
MEASURE mA dc:
Range: 20 mA (24 mA max)
Resolution: 1 μA
Accuracy: ± (0.015 % of reading + 2 counts)
SOURCE / SIMULATE mA dc:
Range: 0 mA to 20 mA (24 mA max)
Resolution: 1 μA
Accuracy: ± (0.015 % of reading + 2 counts)
Source Mode:
Compliance: To 700 Ω at 20 mA
Simulate Mode:
External loop voltage requirement: 24 V nominal,
28 V maximum, 12 V minimum
Loop Power:
Maximum Load: 700 Ω
Percent Display:
25 % to 125 % (4 to 20 mA mode)
Input/Output Protection:
Fused; not replaceable
General Specifications
Maximum voltage between any terminal and earth
ground or between any two terminals:
28 V
Storage Temperature:
30 °C to 60 °C
Operating Temperature:
10 °C to 50 °C
Operating Altitude:
3000 meters maximum
Pollution Degree 2
Temperature
Coefficient:
± 0.005 % of range per °C for temperatures of
10 °C to 18 °C and 28 °C to 50 °C
Relative Humidity:
95 % up to 30 °C
75 % up to 40 °C
45 % up to 50 °C
Vibration:
Random 2 g, 5 to 500 Hz
Shock:
1 meter drop test
Safety Compliance:
Complies with IEC 61010-1-95 CAT I, 28 V; CSA
C22.2 No. 1010-92 NRTL; ANSI/ISA S82.02.01-
1994;
( Directive 94/9/EG and p
NEC 500: Uo = 27.6 V, Io = 96.13 mA, Co = 76 nF,
Lo = 2.5 mH, Ui = 30 V, Ii = 24 mA, Ci = 10 nF, Li =
0 mH
CE:
Complies with EN61010-1 and EN61326
Power Requirements:
Single 9 V battery
(See Table 2 in the User-Replaceable Parts section)
Battery Life (typical):
SOURCE mode: 18 hours; 12 mA into 500 Ω;
MEASURE / SIMULATE mode: 50 hours
Size:
69.85 mm (W) x 142.87 mm (L) x 50.80 mm (H)
[ 2.75 in (W) x 5.625 in (L) x 2.00 in (H) ]
With red holster and Flex-Stand:
76.20 mm (W) x 158.75 mm (L) x 54.61mm (H)
[ 3.00 in (W) x 6.25 in (L) x 2.15 in (H) ]
Weight:
0.28 kg (0.62 lb)
With red holster and Stand:
0.42 kg (0.93 lb)
Safety Information
In this manual a Warning identifies conditions and
actions that pose hazards to the user. A Caution
identifies conditions and actions that may damage the
mA Calibrator or equipment under test. International
symbols used in this manual are identified in the
Symbols section.
Read the entire 707Ex Users Manual and the 707Ex
mA Calibrator CCD (Concept Control Drawing)
before using the mA Calibrator.