Process Calibrators
Required Equipment
25
Snaps
F1
Wh003f.eps
Figure 6. Replacing the Fuse (718 shown)
Required Equipment
The equipment required to perform the verification and calibration procedures in this
manual is identified in Table 33.
Table 33. Required Calibration Equipment
Calibrator
Model
Equipment Minimum Specifications
Recommended
Model
712
DC Calibrator
5 to 3000 Ω
Accuracy: 70 ppm + 0.025 Ω
Fluke 5520A
Reference Multimeter
5 to 3000 Ω
Accuracy: 70 ppm + 0.037 Ω
Fluke 8508A
Four Test Leads 5500A/LEADS
www.GlobalTestSupply.com