11
Quick Start
Unpacking
Quick Start4
Caution: READ SECTION 6 ENTITLED BATH USE before placing the bath in
service. Incorrect handling can damage the bath and void the war ranty.
4.1 Unpacking
Unpack the Micro-Bath carefully and inspect it for any damage that may have
occurred during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
7103 Micro-Bath
L
Transport/Pour Access Lid
L
Probe Basket
L
Stir Bar
L
Power Cord
L
Manual
L
Access Cover, if purchased
L
Setup 4.2
Caution: DO NOT operate the instrument without fluid.
Place the calibrator on a at surface with at least 6 inches of free space around the
instrument. Plug the power cord into a grounded mains outlet. Observe that the
nominal voltage corresponds to that indicated on the back of the calibrator.
Carefully insert the probe basket into the well. Fill the well with the appropriate uid.
The set-point temperature and the number of and size of probes deter mine the uid
level. Be sure to keep the uid level an adequate distance below the top of the well
to prevent overowing the uid when the probes are in serted. For example, placing
200.05 oil at room temperature (25°C) into the bath and heating the unit to 126°C,
causes a 1-inch (2.54 cm) expansion of the uid inside the well.
Keep the uid level at least 1.9 cm (0.75 inches) below the top of the well at all
times. With the probe (probes) in the well ll the tank 3/4 full. Heat to the maximum
temperature of the uid. Slowly ll the well to 2.54 cm (1 inch) be low the top of the
basket at the maximum temperature of the uid, or to the second ring on the basket.