General Operation
Stirring
Pour Point: This represents a handling limit for the uid.
Viscosity: Points shown are at 50 and 10 centistokes viscosity. When viscosity is
greater than 50 centistokes stirring is very poor and the uid is unsatisfactory for bath
applications. Optimum stirring generally occurs at 10 centistokes and below.
Fume Point: A fume hood should be used. This point is very subjective in na ture and
is impacted by individual tolerance to different fumes and smells, how well the bath is
covered, the surface area of the uid in the bath, the size and ventilation of the facility
where the bath is located and other conditions. We as sume the bath is well covered at
this point. This is also subject to company policy.
Flash Point: The point at which ignition may occur. The point shown may be either
the open or closed cup ash point. Refer to the ash point discussion in Section 8.3.8.
Boiling Point: At or near the boiling point of the uid, the temperature stabil ity is
difcult to maintain. Fuming or evaporation is excessive. Large amounts of heater
power may be required because of the heat of vaporization.
Decomposition: The temperature may reach a point at which decomposition of the
uid begins. Further increasing the temperature may accelerate decomposi tion to the
point of danger or impracticality.
Stirring 8.4
Stirring of the bath uid is very important for stable temperature control. The uid
must be mixed well for good temperature uniformity and fast controller response. The
stirrer is adjusted for optimum performance. Table 4 on page 27 shows nominal stirrer
motor settings for several uids.
If the stirrer does not function properly, the instrument will oscillate and not meet
published specications.
Note: If the bath is used with the probe basket removed, stir motor settings need
to be changed so that a small vortex can be seen in the liquid.
Warning: Do not mix water and oil when exceeding temperatures of 90°C
Table 4 Nominal Stirrer Motor Settings With Different Liquids
Liquid Stir Motor Setting Temperature
Distilled Water 12 5°C to 90°C
(41°F to 194°F)
Ethylene Glycol 12 -5°C to 90°C
(25°F to 194°F)
200.05 Oil 15 -30°C to 125°C
(-22°F to 257°F)
200.10 Oil 15 25°C to 125°C
(77°F to 257°F)