7103 Micro-Bath User’s Guide
Scan
34
Un= F
New units selected
Scan 9.3
The scan rate can be set and enabled so that when the set-point is changed the bath
heats or cools at a specied rate (degrees per minute) until it reaches the new set-point.
With the scan disabled the bath heats or cools at the maximum possible rate.
9.3.1 Scan Control
The scan is controlled with the scan on/off function that appears in the main menu
after the set-point function.
Sc=OFF
Scan function off
Press “UP” or “DOWN” to toggle the scan on or off.
Sc=On
Scan function on
Press “SET” to accept the present setting and continue.
2
Accept scan setting
9.3.2 Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set from
.1 to 99.9°C/min. The maximum scan rate however is actually limited by the natural
heating or cooling rate of the instrument. This is often less than 100°C/min, especially
when cooling.
The scan rate function appears in the main menu after the scan control function. The
scan rate units are in degrees C per minute.
Sr= 10.0
Scan rate in°C/min
Press “UP” or “DOWN” to change the scan rate.
Sr= 2.0
New scan rate
Press “SET” to accept the new scan rate and continue.
2
Accept scan rate