7103 Micro-Bath User’s Guide
Interface Commands
46
Table 5 Controller Communications Commands
Command Description Command Format Command Example Returned Returned Example Acceptable Values
Display Temperature
Read current set-point s[etpoint] s set: 999.99 {C or F} set: 150.00 C
Set current set-point to n s[etpoint]=n s=200.00 Instrument Range
Read temperature t[emperature] t t: 999.99 {C or F} t: 55.6 C
Read temperature units u[nits] u u: x u: C
Set temperature units: u[nits]=c/f C or F
Set temperature units to Celsius u[nits]=c u=c
Set temperature units to
Fahrenheit
u[nits]=f u=f
Read scan mode sc[an] sc scan: {ON or OFF} scan:ON
Set scan mode sc[an]=on/off sc=on ON or OFF
Read scan rate sr[ate] sr srat: 99.9 {C or F}/min srat:12.4C/min
Set scan rate sr[ate]=n sr=1.1 .1 to 99.9
Read hold ho[ld] ho hold: open/closed,
99.9 {C or F}
hold: open, 30.5 C
Secondary Menu
Read proportional band setting pr[opband] pr pb: 999.9 pb: 15.9
Set proportional band to n pr[opband]=n pr=8.83 Depends on
Conguration
Read heater power
(duty cycle)
po[wer] po po: 999.9 po: 1.0
Conguration Menu
Operating Parameters Menu
Read stirrer motor speed mo[tor] mo mo: 99 mo: 15
Set stirrer motor speed to n mo[tor]=n mo=16 0 to 40
Read high limit hl hl hl:999 hl:126
Set high limit hl=n hl=90 0–126
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 9 sa: 1
Set serial sampling setting to n
seconds
sa[mple]=n sa=0 0 to 999
Set serial duplex mode: du[plex]=f[ull]/h[alf] FULL or HALF
Set serial duplex mode to full du[plex]=f[ull] du=f
Set serial duplex mode to half du[plex]=h[alf] du=h
Set serial linefeed mode: lf[eed]=on/of[f] ON or OFF
Set serial linefeed mode to on lf[eed]=on lf=on
Set serial linefeed mode to off lf[eed]=of[f] lf=of
Calibration Menu
Read R0 calibration parameter r[0] r r0: 999.999 r0: 100.578
Set R0 calibration parameter to n r[0]=n r=100.324 90 to 110
Read ALPHA calibration
parameter
al[pha] al al: 9.9999999 al: 0.0038573
Set ALPHA calibration parameter
to n
al[pha]=n al=0.0038433 .002 to .005
Read DELTA calibration
parameter
de[lta] de de:9.99999 de: 1.507
Set DELTA calibration parameter de[lta]=n de=1.3742 0–3.0
Read BETA calibration parameter be[ta] be be:99.999 be:03427
Set BETA calibration parameter be[ta]=n be=0.342 –20 to 20
Functions not on menu
Read rmware version number *ver[sion] *ver ver.9999,9.99 ver.7103,2.00
Read structure of all commands h[elp] h list of commands
Read all operating parameters all all list of parameters
Legend: [] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note: When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a carriage return and
linefeed. Then the value is returned as indicated in the RETURNED column.