EasyManua.ls Logo

Fluke 74X Series - DC Voltage Output; DC Current Output; Resistance Sourcing

Fluke 74X Series
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
5
DC Voltage Output
Range Resolution % of Output + % of Full Scale
1 Year 2 Year
110 mV 1 µV 0.01% + 0.005% 0.015% + 0.005%
1.1 V 10 µV 0.01% + 0.005% 0.015% + 0.005%
15 V 100 µV 0.01% + 0.005% 0.015% + 0.005%
Temperature Coefficient: (0.001% of output + 0.001% of f.s.)/ °C in the ranges -10 to 18 °C and 28 to 50
°C
Maximum Output Current: 10 mA
Loading: (0.001% f.s. + 1 µV)/ mA
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc
DC Current Output
Range/Mode Resolution % of Output + % of Full Scale
1 Year 2 Year
22 mA/ Source mA 1 µA 0.01% + 0.015% 0.02% + 0.015%
22 mA/ Simulate Transmitter
(Current Sink)
1 µA 0.02% + 0.03% 0.02% + 0.03%
Maximum Burden Voltage: 24V
Temperature Coefficient: (0.003% of output + 0.003% of f.s.)/ °C in the ranges -10 to 18 °C and 28 to
50 °C
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc
Specification applies for currents between 2 mA and 22 mA. For current below 2 mA, typical accuracy is
0.15% of full scale.
Resistance Sourcing
Range Resolution % of Output + ohms
1 Year 2 Year
11.000 1 m 0.01% + 0.02 0.02% + 0.02
110.00 10 m 0.01% + 0.04 0.02% + 0.04
1.1000 k 100 m 0.02% + 0. 5 0.03% + 0.5
11.000 k 1 0.03% + 5 0.04% + 5
Temperature Coefficient: (0.01% of f.s.)/ °C in the ranges -10 to 18 °C and 28 to 50 °C
Maximum and Minimum Current through Source Resistance:
11 Range: 3 mA dc max, 0.1 mA dc min
110 Range: 3 mA dc max, 0.1 mA dc min
1.1 k Range: 3 mA dc max, 0.01 mA dc min
11 k Range: 1 mA dc max, 0.01 mA dc min
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc

Related product manuals