74X Series
Calibration Manual
16
Resistance Measurement
Proceed as follows to verify the Resistance Measurement function:
1. Connect the UUT to the 5500A as shown in Figure 4. Use a four-wire connection at
the 5500A, transitioning to two wires at the UUT, and turn Two-Wire Compensation
on.
2. Set the UUT to the Resistance Measurement function.
3. Press [RANGE] on the UUT to lock on the 11 range.
4. Set the 5500A to the first test point in Table 6.
5. Observe that the reading on the UUT is within the range shown under the appropriate
column.
6. Continue through the test points, being careful to lock the UUT on the specified
range.
7. When you are finished, set the 5500A to [STANDBY].
TC
mA mA
V
V
RTD
RTD
SOURCE
300V
MAX
30V
MAX
30V
MAX
30V
MAX
MEAS
CAT
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
FLUKE 74X FLUKE 5500A CALIBRATOR
2-WIRE COMP ON
gb05f.eps
Figure 4. Resistance Measurement Verification Connections
Table 6. Resistance Measurement Verification Points
UUT
Range
Input
Minimum
1 Year
Maximum
1 Year
Minimum
2 Year
Maximum
2 Year
11 ohm 0.0 ohm -0.050 0.050 -0.050 0.050
11 ohm 10 ohm 9.945 10.055 9.943 10.058
110 ohm 0.0 ohm -0.05 0.05 -0.05 0.05
110 ohm 100 ohm 99.90 100.10 99.88 100.13
1100 ohm 0.0 ohm -0.5 0.5 -0.5 0.5
1100 ohm 1000 ohm 999.0 1001.0 998.8 1001.3
11 kohm 0.0 ohm -0.010 0.010 -0.010 0.010
11 kohm 10 kohm 9.980 10.020 9.980 10.020