Performance Verification Tests
13
Table 3. DC Volts Measurement Verification Points
UUT
Range
Input
DCV
Minimum
1 Year
Maximum
1 Year
Minimum
2 Year
Maximum
2 Year
11
mV 0 mV -0.017 0.017 -0.017 0.017
11
mV 100 mV 99.959 100.042 99.934 100.067
11
mV -100 mV -100.042 -99.959 -100.067 -99.934
1.1 V 0 mV -0.00006 0.00006 -0.00006 0.00006
1.1 V 1 V 0.99970 1.00031 0.99945 1.00056
1.1 V -1 V -1.00031 -0.99970 -1.00056 -0.99945
11 V 0 V -0.0006 0.0006 -0.0006 0.0006
11 V 10 V 9.9970 10.0031 9.9945 10.0056
11 V -10 V -10.0031 -9.9970 -10.0056 -9.9945
11
V 0 mV -0.006 0.006 -0.006 0.006
11
V 100 V 99.945 100.056 99.895 100.106
11
V -100 V -100.056 -99.945 -100.106 -99.895
30
V 0 mV -0.02 0.02 -0.02 0.02
30
V 300 V 299.84 300.17 299.69 300.32
30
V -300 V -300.17 -299.84 -300.32 -299.69