EasyManua.ls Logo

Fluke 74X Series - Page 21

Fluke 74X Series
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification Tests
13
Table 3. DC Volts Measurement Verification Points
UUT
Range
Input
DCV
Minimum
1 Year
Maximum
1 Year
Minimum
2 Year
Maximum
2 Year
11
0
mV 0 mV -0.017 0.017 -0.017 0.017
11
0
mV 100 mV 99.959 100.042 99.934 100.067
11
0
mV -100 mV -100.042 -99.959 -100.067 -99.934
1.1 V 0 mV -0.00006 0.00006 -0.00006 0.00006
1.1 V 1 V 0.99970 1.00031 0.99945 1.00056
1.1 V -1 V -1.00031 -0.99970 -1.00056 -0.99945
11 V 0 V -0.0006 0.0006 -0.0006 0.0006
11 V 10 V 9.9970 10.0031 9.9945 10.0056
11 V -10 V -10.0031 -9.9970 -10.0056 -9.9945
11
0
V 0 mV -0.006 0.006 -0.006 0.006
11
0
V 100 V 99.945 100.056 99.895 100.106
11
0
V -100 V -100.056 -99.945 -100.106 -99.895
30
0
V 0 mV -0.02 0.02 -0.02 0.02
30
0
V 300 V 299.84 300.17 299.69 300.32
30
0
V -300 V -300.17 -299.84 -300.32 -299.69

Related product manuals