753/754
Users Manual
70
9. Push for MEASURE/SOURCE mode. The display
changes to the screen shown in Figure 28.
gks42s.bmp
Figure 28. Process Instrument Calibration Screen
10. Push the As Found softkey, followed by the
Instrument selection ().
The display changes to the screen shown in Figure 29.
gks44s.bmp
Figure 29. Process Instrument Calibration Screen 2
11. Record values for 0% and 100% of 4.0 mA and
20.0 mA, in that sequence. Set Tolerance to 0.5% of
span.