EasyManua.ls Logo

Fluke 78 - Ohms Test; Continuity Test

Fluke 78
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
Performance Tests
3
3-11
3-19. Ohms Test
1. Set the Fluke 78 rotary switch to the e function.
2. Connect the Ohms Calibrator or Decade Resistor to the z and COM input
terminals of the meter.
3. Referring to Table 3-7, set the Decade Resistor or Ohms Calibrator to the resistance
value indicated in steps 1 through 7. Verify that the display reading is within the
limits shown in the table.
Table 3-7. Resistance Test
Step Range Input
Resistance
Display Reading
1 400 J short 000.0 to 000.2 J
Decades of 1:
2 400 J 100 J 99.3 to 100.7 J (plus 0 reading)
34 kJ 1 kJ .994 to 1.006 kJ
4 40 kJ 10 kJ 9.94 to 10.06 kJ
5 400 kJ 100 kJ 99.4 to 100.6 kJ
64 MJ 1 MJ .994 to 1.006 MJ
7 40 MJ 10 MJ 9.87 to 10.13 MJ
Decades of 1.9:
2 400 J 190 J 188.8 to 191.2 J(plus 0 reading)
34 kJ 1.9 kJ 1.889 to 1.911 kJ
4 40 kJ 19 kJ 18.89 to 19.11 kJ
5 400 kJ 190 kJ 188.9 to 191.1 kJ
64 MJ 1.9 MJ 1.889 to 1.911 MJ
7 40 MJ 19 MJ 18.78 to 19.22 MJ
Decades of 3.5:
2 400 J 350 J 348.0 to 352.0 J (plus 0
reading)
34 kJ 3.5 kJ 3.481 to 3.519 kJ
4 40 kJ 35 kJ 34.81 to 35.19 kJ
5 400 kJ 350 kJ 348.1 to 351.9 kJ
64 MJ 3.5 MJ 3.481 to 3.519 MJ
7 40 MJ 35 MJ 34.62 to 35.38 MJ
3-20. Continuity Test
1. Set the Fluke 78 rotary switch to w and press the pushbutton for 2 seconds to put
the meter in the continuity test function.
2. Referring to Table 3-8, apply inputs as indicated. Verify that the Fluke 78 display
and beeper indicate as shown in the table.
Table 3-8. Continuity Test
Step Range Input Display Reading
1 400 J open OL
2 400 J 30 J tone

Other manuals for Fluke 78

Related product manuals