High-Pressure Modular Controller
Controller Settings (Setup Menu)
21
Remote Port Menu
Use the Remote Port menu to change or view the USB, GPIB, RS-232, and
Ethernet port settings explained in Table 6.
Table 6. Remote Port Menu
Tab Description
USB Setup
Selects whether the remote interface (Remote IF) is from a Computer or Terminal.
Specifies the End of Line character (EOL) as Carriage Return (CR), Line Feed (LF),
or Carriage Return and Line Feed (CRLF).
RS-232 Setup
Use this menu to specify the RS-232 communication parameters. The editable
parameters are:
Data Bits
Stop Bits
Flow Control
Parity
Baud
EOL
Remote IF
Set EOF
Touch Defaults to reset the RS-232 port to its default values.
Emulation
Mode
From this menu, and with remote commands, the Product can emulate a variety of
different instruments. Refer to the Remote Programmers Guide on the Fluke
Calibration website for a list of emulated products.
GPIB
(IEEE-488)
Address
Specify the GPIB Address from this tab. Touch the tab and use the numeric keypad
on the Product or PC to change the address.
Ethernet Setup
Use this menu and the numeric keypad or PC keypad to specify or edit these
parameters:
Host Name
IP Address
Gateway
Subnet Mask
DHCP (ON or OFF)
Remote IF (interface)
Port