867B/863
Users Manual
2-10
Using the Ohms Meter Mode Softkeys
Rel
/
nS
2
(
/
) enables the Continuity beeper. The GMM
beeps each time a low resistance continuity
measurement is encountered. For example,
beginning in the 300Ω range, the beeper goes on
when the resistance passes below 32Ω and off when
the resistance rises above 132Ω. You can select a
different range after activating the Continuity beeper;
each range uses different beeper on and off points.
For a list of these on and off points, see Continuity
Specifications in Chapter 8 of this manual.
3
(nS) begins conductance (nanosiemens “nS”)
readings in the primary display. Measurement
ranges are 300 nS and 3000 nS. (Megohm-only
readings continue in the secondary display.) Press
3
again to return to ohms (Ω) primary readings
(all ranges). Selecting conductance (nS)
automatically cancels active Min Max, “Peak Hold”,
Touch Hold, Continuity, and “Rel, and resets the
Trend plot. Frequency readings are not available in
conductance or ohms.