EasyManua.ls Logo

Fluke 863 - Operational Test; Self Test

Fluke 863
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Maintenance
Operational Test
7
7-5
Operational Test
Use the following procedure to test the basic
operability of your GMM test tool:
1. With no inputs connected, turn the GMM test tool
on by rotating the selector to VAC (
v
). Combo or
Meter display mode should appear, with a
random display in the mV range.
2. Rotate the selector to VDC (
E
) Observe a similar
primary display in the mV range.
3. Rotate the selector to Ohms (
R
). Observe a
display of “OL”.
4. Connect the test leads to
V
and
C
.
5. Short the test lead tips and observe a display of
approximately 0.250 ohms.
6. Press
2
(
/
), short the test leads again, and
listen for a beep.
For a complete performance test of GMM test tool
functions and specifications, refer to the 863/867B
Service Manual (PN 689312).
Self Test
Begin an internal diagnostic routine by selecting SET
UP, wait two seconds, press
d2
. The display
now presents the following information:
1. A test pattern.
2. An error message (if a problem is encountered
during the test). The GMM test tool Model and
serial numbers.
3. The calibration status.
Generally, you cannot use the GMM test tool keys
during Self Test. (You can press
1
to exit the
diagnostic routine and return to the SET UP
Softkeys.)

Table of Contents

Other manuals for Fluke 863

Related product manuals