Accuracy
Function Range Resolution 45-1 kHz 20-45 Hz 1-20 kHz 20 kHz-100 kHz
AC µA 500.00 µA 0.01 µA 0.75 % + 20 1 % + 20 0.75 % + 20 6 % + 40
5,000.0 µA0.1 µA 0.75 % + 5 1% + 5 0.75 % + 10 2 % + 40
AC mA 50.000 mA 0.001 mA 0.75 % + 20 1% + 20 0.75 % + 20 9 % + 40
400.00 mA
1
0.01 mA 0.75 % + 5 1% + 5 1.5 % + 10 4 % + 40
AC A 5.0000 A 0.0001 A 1.5 % + 20 1.5% + 20 6 % + 40 Not specified
10.000 A
2
0.001 A 1.5 % + 5 1.5% + 5 5 % + 10 Not specified
1. 500.00 mA overload for 30 seconds maximum.
2. 10 A continuous up to 35 °C, less than 10 minutes -35 °C to 55
°C. 20 A overload for 30 seconds maximum.