EasyManua.ls Logo

Fluke 87 Series IV - Page 91

Fluke 87 Series IV
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Detailed Accuracy Specifications
7
7-7
Accuracy Accuracy Dual Display AC or AC+DC
6
Function Range Resolution DC 20 - 45 Hz 45 Hz - 1 kHz 1 kHz- 20 kHz
DC mV 50.000 mV 0.001 mV 0.1% + 20
500.00 mV 0.01 mV 0.03 % + 2
4
2 % + 80 0.6 % + 40
3000.0 mV
7
0.1 mV 0.025 % + 5 (89-IV)
0.025 % + 10 (87-IV)
2 % + 80 0.6 % + 40 6 % + 40
3
DC V 5.0000 V 0.0001 V 0.025 % + 10 2 % + 80 0.5 % + 40
50.000 V 0.001 V 0.03 % + 3 2 % + 80 0.5 % + 40
500.00 V 0.01 V 0.1 % + 2 2 % + 80 0.5 % + 40 Not specified
1000.0 V 0.1 V 0.1 % + 2 2 % + 80 0.5 % + 40 Not specified
DC µA 500.00 µA 0.01 µA 0.25 % + 20
5
7 % + 10 7 % + 10 9 % + 40
5,000 µA0.1 µA 0.25 % + 2 1 % + 10 0.75 % + 10 2 % + 40
DC mA 50.000 mA 0.001 mA 0.15 % + 10 1 % + 10 0.75 % + 10 2 % + 40
400.00 mA
1
0.01 mA 0.15 % + 2 1.5 % + 10 1.5 % + 10 3 % + 40
DC A 5.0000 A 0.0001 A 0.5 %+ 10 7 % + 20 7 % + 20 12 % + 40
10.000 A
2
0.001 A 0.5 %+ 2 1.5 % + 10 1.5 % + 10 3 % + 40
1. 500.00 mA for 30 seconds maximum.
2.
10 A continuous up to 35 °C, less than 10 minutes -35 °C to 55 °C. 20 A overload for 30 seconds maximum.
3. DC 5 V range, -3 dB typical @ 10 kHz.
4. In RF field of 3 V/m, add 100 counts from 100 to 120 MHz, 60 counts from 270 to 300 MHz, 40 counts from 320 to 335 MHz.
5. In RF filed of 3 V/m, add 200 counts from 80 to 95 MHz.
6. See AC conversions notes for AC mV and V.
7. 1100.0 mV AC or AC+DC.

Table of Contents

Related product manuals