Multimeters 
  Making Measurements   
 15 
Zero Input Behavior of True RMS Meters (87) 
True Rms Meters accurately measure distorted 
waveforms, but when the input leads are shorted together 
in the AC functions, the meter displays a residual reading 
between 1 and 30 counts. When the test leads are open, 
the display readings may fluctuate due to interference. 
These offset readings are normal. They do not affect the 
Meter’s AC measurement accuracy over the specified 
measurement ranges. 
Unspecified input levels are: 
•  AC voltage: below 3 % of 600 mV AC, or 18 mV AC 
•  AC current: below 3 % of 60 mA AC, or 1.8 mA AC 
•  AC current: below 3 % of 600 μA AC, or 18 μA AC 
Low Pass Filter (87) 
The 87 is equipped with an ac low pass filter. When 
measuring ac voltage or ac frequency, press A to 
activate the low pass filter mode (K). The Meter continues 
measuring in the chosen ac mode, but now the signal 
diverts through a filter that blocks unwanted voltages above 
1 kHz, refer to Figure 3. The lower frequency voltages pass 
with reduced accuracy to the measurement below 1 kHz. 
The low pass filter can improve measurement performance 
on composite sine waves that are typically generated by 
inverters and variable frequency motor drives. 
XW Warning 
To avoid possible electric shock or personal 
injury, do not use the Low Pass Filter option 
to verify the presence of hazardous voltages. 
Voltages greater than what is indicated may 
be present. First, make a voltage measure-
ment without the filter to detect the possible 
presence of hazardous voltage. Then, select 
the filter function. 
Note 
In Low Pass Mode, the Meter goes to manual 
mode. Select ranges by pressing the RANGE 
button. Autoranging is not available in Low Pass 
Mode. 
1 kHz
100 Hz
 
aom11f.eps 
Figure 3. Low Pass Filter