80 Series V 
Users Manual 
24 
Measuring AC or DC Current 
XW Warning 
To avoid possible electric shock or personal 
injury, never attempt an in-circuit current 
measurement where the open-circuit 
potential to earth is greater than 1000 V. You 
may damage the Meter or be injured if the 
fuse blows during such a measurement. 
W Caution 
To avoid possible damage to the Meter or to 
the equipment under test: 
•  Check the Meter's fuses before 
measuring current. 
•  Use the proper terminals, function, and 
range for all measurements. 
•  Never place the probes across (in 
parallel with) any circuit or component 
when the leads are plugged into the 
current terminals. 
To measure current, you must break the circuit under 
test, then place the Meter in series with the circuit. 
The Meter's current ranges are 600.0 μA, 6000 μA, 
60.00 mA, 400.0 mA, 6000 mA, and 10 A. AC current is 
displayed as an rms value. 
To measure current, refer to Figure 8 and proceed as 
follows: 
1.  Turn off power to the circuit. Discharge all high-
voltage capacitors. 
2.  Insert the black lead into the COM terminal. For 
currents between 6 mA and 400 mA, insert the red 
lead into the mA/μA terminal. For currents above 
400 mA, insert the red lead into the A terminal. 
Note 
To avoid blowing the Meter's 400 mA fuse, use 
the mA/
μ
A terminal only if you are sure the 
current is less than 400 mA continuously or less 
than 600 mA for 18 hours or less.