Multimeters 
  HiRes Mode (Model 87)   
 31 
Zoom Mode (Power Up Option Only) 
To use the Rel Zoom Bar Graph: 
1. Hold down F while turning the Meter on. The 
display reads “2rEL”. 
2.  Select the relative mode by pressing F 
again. 
3.  The center of the bar graph now represents zero 
and the sensitivity of the bar graph increases by 
a factor of 10. Measured values more negative 
than the stored reference activate segments to 
the left of center; values more positive activate 
segments to the right of center. 
Uses for the Zoom Mode  
The relative mode, combined with the increased 
sensitivity of the bar graph's zoom mode, helps you make 
fast and accurate zero and peak adjustments. 
For zero adjustments, set the Meter to the desired 
function, short the test leads together, press 
F; then 
connect the leads to the circuit under test. Adjust the 
circuit's variable component until the display reads zero. 
Only the center segment on the zoom bar graph is lit.  
For peak adjustments, set the Meter to the desired 
function, connect the leads to the circuit under test; then 
press 
F. The display reads zero. As you adjust for a 
positive or negative peak, the bar graph length increases 
to the right or left of zero. If an overange symbol lights (< 
>), press 
F twice to set a new reference; then 
continue with the adjustment. 
HiRes Mode (Model 87) 
On a Model 87 Meter, pressing H for one second causes 
the Meter to enter the high-resolution (HiRes), 4-1/2 digit 
mode. Readings are displayed at 10 times the normal 
resolution with a maximum display of 19,999 counts. The 
HiRes mode works in all modes except capacitance, 
frequency counter functions, temperature, and the 250 μs 
(peak) MIN MAX modes. 
To return to the 3-1/2 digit mode, press H again for one 
second.