Remote Operation
Supported SCPI Commands
47
Examples PER:APER 0.1 Sets the period measurement gate time to 100
ms.
Query PER:APER? Returns the gate time for the period function.
[SENSe:]CURRent:AC:BANDwidth(?) <n>
Description Sets the appropriate filter for the frequency specified by <n>. This setting
is stored in volatile memory.
Parameters <n> = 3 Selects slow filter
20 (default) Selects medium filter
200 Selects fast filter
MIN Selects slow filter
MAX Selects fast filter
Examples CURR:AC:BAND 200 Sets the filter for ac current to fast.
Query CURR:AC:BAND? Returns the filter selection.
[SENSe:]VOLTage:AC:BANDwidth(?) <n>
Description Sets the appropriate filter for the frequency specified by <n>. This setting
is stored in volatile memory.
Parameters <n> = 3 Selects slow filter
20 (default) Selects medium filter
200 Selects fast filter
MIN Selects slow filter
MAX Selects fast filter
Examples VOLT:AC:BAND 20 Sets the filter for ac voltage to medium.
Query VOLT:AC:BAND? Returns the filter selection.
[SENSe:]DETector:BANDwidth(?) <n>
Description Sets the appropriate filter for the frequency specified by <n>. This setting
is stored in volatile memory and sets the filter for all functions that use an
ac filter.
Parameters <n> = 3 Selects slow filter
20 Selects medium filter
200 Selects fast filter
MIN Selects slow filter
MAX Selects fast filter
Examples DET:BAND 20 Sets the filter to medium.
Query DET:BAND? Returns the filter selection.
DET:BAND? MAX Returns +2.00000000E+02 to indicate the fast
filter.
[SENSe:]FILTer[:DC][:STATe](?) <n>
Description Activates or deactivates the 3-pole analog filter to improve noise immunity
for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the analog dc filter off.
ON or 1 Turns the analog dc filter on.