EasyManua.ls Logo

Fluke 902 - General Specifications

Fluke 902
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
902
Calibration Information
4
General Specifications
Operating Temperature -10 °C to +50 °C
Storage Temperature -40 °C to +60 °C
Operating Humidity
Non condensing (< 10 °C)
90 % RH (10 °C to 30 °C)
75 % RH (30 °C to 40 °C)
45 % RH (40 °C to 50 °C)
(Without Condensation)
Operating Altitude 2500 meters above mean sea level
Storage Altitude 12,000 meters above mean sea level
IP Rating IP 30 per IEC 60529
Vibration Requirements MIL-PRF-28800F Class 2 random vibration
EMI, RFI, EMC
EMI: instrument unspecified for use in EMC field 0.5 V / Meter
EMC: Meets all applicable
requirements in EN61326-1
Temperature
Coefficients
0.1 x (specified accuracy)/ °C
(<18 °C or >28 °C)
Size (H X W X L)
9.1 x 3.8 x 1.7 inches
(240 x 80 x 40 mm)
Weight
1.1 lb
(310 g)
Design Standards and
Compliance
EN/IEC 61010-1, EN/IEC 61010-2-032 and relevant U.S. and Canadian Standards
Agency Approvals
P ) ; s
N10140
Measurement Category
600 V, CAT III
CAT III equipment is designed to protect against transients in equipment in fixed-equipment
installations, such as distribution panels, feeders and short branch circuits, and lighting
systems in large buildings.
Power Requirements Two AA Batteries, NEDA 15 A, IEC LR6