HVAC Clamp Meter
Performance Tests
9
Table 2. Performance Tests
Meter Response
Test
(Switch Position)
5520A Output
Lower Limit Upper Limit
600 V/60 Hz 593.5 606.5
V
AC Voltage
10 V/60 Hz 9.4 10.6
600 V dc 593.5 606.5
0 V dc -0.5 0.5
V
DC Voltage
-600 V dc -606.5 -593.5
P
Continuity
25 Beeper on
9500 9353 9647
1500 1473 1527
950 935.3 964.7
P
Resistance
0 -0.5 0.5
200 µA dc 197.5 202.5
10 µA dc 9.4 10.6
0 µA dc -0.5 0.5
G
DC Current
-200 µA dc -202.5 -197.5
-40 °C -43.5 -36.5
0 °C -0.8 0.8
**F
Temperature
400 °C 395.2 404.8
1 µF 0.8 1.2
90 µF 88.1 91.9
200 µF 195 205
L
Capacitance
1 mF 979 1021
0.4 A, 60 Hz 19.1 20.9
12 A, 60 Hz 587.5 612.5
*K
AC Current
12 A, 50Hz 587.5 612.5
*Test using the 5500A Coil
**Use K-Type TC Extension Wire with 80CK, at both ends, for connection between 5520A and 80AK/Meter