21
Controller Operation
Temperature Set-point
To change to another set-point memory press the up or down arrow.
4 50. New set-point memory 4 location, 50°C
Press “SET” to accept the new selection and access the set-point value. Press “EXIT”
to continue and to ignore any changes.
S
Accept selected set-point memory
Set-point Value6.2.2
The set-point value may be adjusted after selecting the set-point memory and pressing
“SET”.
4 50. Set-point 4 value in°C
If the set-point value does not need to be changed, press and hold “EXIT” to resume
displaying the well temperature. To change the set-point value, press “SET” and then
press the up or down arrow.
60. New set-point value
When the desired set-point value is reached, press “SET” to accept the new value and
access the temperature scale units selection. If “EXIT” is pressed, any changes made
to the set-point are ignored.
S
Accept new set-point value
Temperature Scale Units6.2.3
The temperature scale units of the controller maybe set by the user to degrees Celsius
(°C) or Fahrenheit (°F). The units are used in displaying the well temperature, set-
point, proportional band, and high limit.
Press “SET” after adjusting the set-point value to change display units.
Un= C Scale units currently selected
Press the up or down arrow to change the units.
Un= F New units selected
Press “SET” to accept the new units or “EXIT” to cancel.
Note: The temperature scale units may also be changed by pressing “SET” and u
when the temperature is displayed.