39
Calibration Procedure
Calibration Procedure
Press “SET” and “EXIT” at the same time. Press “SET” until 5. PAR is displayed.
and then press u until CAL is displayed.
Press “SET” five times to enter the menu.6.
Press “SET” and 7. u or d until the correct numerical setting is displayed. Press
“SET” to accept the new value.
Repeat step 3. for ALPHA and DELTA.8.
Press “EXIT” to show the displayed temperature.9.
Accuracy and Repeatability9.2.3
Check the accuracy of the dry-well at various points over the calibration range. If dry-
well does not pass specication at all set-points, repeat the Calibration Procedure.