EasyManua.ls Logo

Fluke a3002 FC - Specifications

Fluke a3002 FC
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
a3002 FC
Calibration Manual
4
Specifications
Range
A dc ................................................................. 1000 A
A ac ................................................................. 600.0 A
Resolution
A dc ................................................................. 0.1 A
A ac ................................................................. 0.1 A
Accuracy
A dc ................................................................. (0.5% + 3 counts)
A ac ................................................................. (1.0% + 3 counts) from 45 Hz to 1000 Hz
LCD w/Backlight .................................................... 3 ½ digits
Log Rate/Interval ................................................... 1 second to 1 hour adjustable by PC, default, 1 minute
Battery Type .......................................................... 2 AA, NEDA 15 A, IEC LR6
Battery Life ............................................................ 400 hours
Memory .................................................................. Record a maximum of 65,000 readings
Radio Frequency Communications .................... 2.4 GHz ISM Band
Radio Frequency Communication Range .......... 20 m (65.61 ft)
Radio Frequency Certification ............................. FCC: T68-DMFBLE; IC: 6627A-DMFBLE
Operating Temperature ........................................ -10 °C to +50 °C (14 °F to 122 °F)
Storage Temperature ............................................ -40 °C to +60 °C (-40 °F to 140 °F)
Operating Humidity ............................................... 90 % from 0 °C to 35 °C, 45 % to 40 °C, 45 % to 50 °C
(90 % from 32 °F to 95 °F, 45 % to 104 °F, 45 % to 122 °F)
Operating Altitude ................................................. 2,000 m (6,562 ft)
Storage Altitude .................................................... 12,000 m (39,370 ft)
Ingress Protection (IP) rating............................... IP42
Safety ..................................................................... IEC 61010-1, Pollution Degree 2
Electromagnetic Environment ............................. IEC 61236-1, Portable
Electromagnetic Compatibility ............................ Radio Frequency Emissions, IEC CISPR 11: Group 1, Class A.
Group 1 have intentionally generated and/or use conductively coupled
radio-frequency energy which is necessary for the internal functioning
of the equipment itself.
Class A equipment is suitable for use in non-domestic locations and/or
directly connected to a low-voltage power supply network. Class A
equipment may have potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted as well as
radiated disturbances.
Applies to use in Korea only. .............................. Class A Equipment (Industrial Broadcasting & Communication
Equipment)
[1]
[1] This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and is not to be
used in homes.
Temperature Coefficient ....................................... Add 0.1 X (specified accuracy)/ °C (<18 °C or >28 °C)
Add 0.1 X (specified accuracy)/ °F (<64.4 °F or >82.4 °F)
Size ......................................................................... 160 mm x 66 mm x 38 mm (6.3 in x 2.6 in x 1.5 in)
Weight .................................................................... 0.255 kg (9 oz)