a3002 FC
Calibration Manual
6
Performance Tests
Warning
To prevent possible electrical shock, fire, or personal injury,
do not perform the performance test procedures unless the
Product is fully assembled.
The performance tests verify the full operation of the Product and measure the
accuracy of each function against Product specifications. If the Product fails a
part of the test, calibration adjustment and/or repair is necessary. See
“Calibration Adjustment”.
Test the Display
To verify that all segments of the display function:
1. With the Product OFF, push and hold .
2. Push while you keep pushed until all of the display segments are
shown. See Figure 1.
hby002.eps
Figure 1. All Segments of the Display
If segments of the display are missing, repair is necessary. See “Contact
Fluke”.