Wireless AC/DC Current Clamp 
  Calibration Adjustment 
11 
Calibration Adjustment 
The Product features closed-case calibration adjustment and uses known 
reference sources. The Product measures the applied reference source, 
calculates correction factors, and stores the correction factors in nonvolatile 
memory. 
Should the Product fail any of the performance tests, do the calibration 
adjustment procedure. 
When “C-01” is shown on the display, apply the correct input signal shown in 
Table 4 to the Product. Then push  to confirm the calibration step. If the input 
signal does not satisfy the calibration requirement, “Err” is shown. If the signal is 
not stable, it can be necessary to push  several times to confirm the calibration.  
After confirmation, the Product goes to the subsequent calibration step. 
Note 
After you push , wait until the calibration step number advances 
before you change the calibrator source. Some adjustment steps 
can take several seconds to execute before the Product goes to the 
subsequent step. 
Set the Calibrator to Standby after you complete adjustment of each 
function.  
Input each signal to the Product in the sequence shown in Table 4. When the last 
calibration point is recorded, “End” shows on the display.  
Note 
While the calibration adjustment points are shown in Table 4, the 
Product also can show the necessary inputs. For each step, push 
 to see the necessary current input and then push  to see the 
necessary frequency input. 
Table 4. Calibration Adjustment 
Calibration Step  Calibrator Output Signal 
C-01  6.0 mV, 55 Hz 
C-02  60.0 mV, 55 Hz 
C-03  600.0 mV, 55 Hz 
C-04  1000.0 mV, 0 Hz