Electrical Safety Analyzer
Safety Information
3
 Warning
To prevent possible electrical shock, fire, or
personal injury, follow these guidelines:
Read all safety information before you use
the Product.
Use the Product only as specified, or the
protection supplied by the Product can be
compromised.
Do not connect to live output terminals.
The Product can supply voltages that can
cause death. Standby mode is not
sufficient to prevent electrical shock.
Do not apply more than the rated voltage,
between the terminals or between each
terminal and earth ground.
Limit operation to the specified
measurement category, voltage, or
amperage ratings.
Use Product-approved measurement
category (CAT), voltage, and amperage
rated accessories (probes, test leads, and
adapters) for all measurements.
Measure a known voltage first to make
sure that the Product operates correctly.
Use the correct terminals, function, and
range for measurements.
Do not touch voltages > 30 V ac rms, 42 V
ac peak, or 60 V dc.
Do not use the Product around explosive
gas, vapor, or in damp or wet
environments.
Do not use the Product if it is damaged.
Disable the Product if it is damaged, by
removing the power cord and fuses.
Do not use the Product if it operates
incorrectly.
Examine the case before you use the
Product. Look for cracks or missing
plastic. Carefully look at the insulation
around the terminals.
Do not use test leads if they are damaged.
Examine the test leads for damaged
insulation, exposed metal, or if the wear
indicator shows. Check test lead
continuity.
Use this Product indoors only.