Electrical Safety Analyzer
Maintenance
25
F1 - F2
hpf111.eps
Figure 14. Fuse Access
Cleaning the Analyzer
 Warning
To avoid electric shock, do not clean the
Analyzer plugged into mains or attached to a
DUT.
Caution
Do not pour fluid onto the Analyzer surface;
fluid seepage into the electrical circuitry may
cause the Analyzer to fail.
Caution
Do not use spray cleaners on the Analyzer;
such action may force cleaning fluid into the
Analyzer and damage electronic components.
Clean the Analyzer occasionally utilizing a damp cloth and
mild detergent. Take care to prevent the entrance of
liquids.
Wipe down the adapter cables with the same care. Inspect
them for damage to and deterioration of the insulation.
Check the connections for integrity before each use.