Electrical Safety Analyzer
Connect the DUT to the Analyzer
9
Connect the DUT to the Analyzer
A Device Under Test (DUT) can be connected in a number
of different ways depending on the device and the number
of connections needed for a full electrical safety test.
Figure 4 shows a DUT connected to the test receptacle
and a separate connection to the DUT’s enclosure or
protective earth ground.
Warning
To prevent possible electrical shock, fire, or
personal injury, follow these guidelines:
Do not touch exposed metal on banana
plugs, they can have voltages that could
cause death.
Remove circuit power before you connect
the Product in the circuit when you
measure current. Connect the Product in
series with the circuit.
Connect an approved three-conductor
mains power cord to a grounded power
outlet.
Do not put the Product where access to the
mains power cord is blocked.
Do not put metal objects into connectors.