Features and Connections
See Table 1 for a list of features and connections.
DC Current Measurement
See Table 2 for instructions on how to make a dc current measurement.
Note
Maximum 400 A dc rms (i410) or 1000 A dc rms (i1010).
Table 1. Features and Connections
Item Description
Jaw Lever
Jaw Centering Marks
ON Indicator
ON/OFF Switch
Zero Adjust
Battery Access
Output Cable
Tactile Barrier
Voltmeter
[1]
[1] Minimum voltmeter requirements:
Accepts safety-shrouded banana plugs
Can display 1 mV (0.1 mV preferred)
Accuracy 0.75 %
Input impedance 1 MΩ, ≤100 pF
Table 2. DC Current Measurement
Step Description
Connect to voltmeter.
Select mV dc.
Set to ON.
Adjust ZERO (jaws empty).
Clamp and center around the
conductor.
Read voltmeter
(1 mV = 1 A).
CLAMP METER
i410
ZEROA
600V
CAT
600V
600A
CLAMP METER
i410
ZEROA
CAT
600V
600A
600V
mV
76
TRUE RMS MULTIMETER
1
2
3
4
6
5
ZEROA
CLAMP METER
i410
ZERO
A
CAT
600V
600A
600V
V
76
TRUE RMS MULTIMETER
1
2
8
3
4
5
6
7
9