Motor Drive Analyzer
Specifications
29
Specifications
DC Voltage (V DC)
Maximum voltage with 10:1 or 100:1 probe..............1000 V
Maximum resolution with 10:1 or 100:1 probe..........1 mV
Full scale reading...................................................... 999 counts
Accuracy at 4 s to 10 us/div......................................±(3 % + 6 counts)
AC voltage (V AC)
Maximum voltage with 10:1 or 100:1 probe..............1000 V
Maximum resolution with 10:1 or 100:1 probe..........1 mV
Full scale reading...................................................... 999 counts
50 Hz ........................................................................±(3 % + 10 counts) -0.6 %
60 Hz ........................................................................±(3 % + 10 counts) -0.4 %
60 Hz to 20 kHz ........................................................±(4 % + 15 counts)
20 kHz to 1 MHz .......................................................±(6 % + 20 counts)
1 MHz to 25 MHz ......................................................±(10 % + 20 counts)
True-rms voltage (V AC+DC)
Maximum voltage with 10:1 or 100:1 probe..............1000 V
Maximum resolution with 10:1 or 100:1 probe..........1 mV
Full scale reading...................................................... 1100 counts
DC to 60 Hz .............................................................. ±(3 % + 10 counts)
60 Hz to 20 kHz ........................................................±(4 % + 15 counts)
20 kHz to 1 MHz .......................................................±(6 % + 20 counts)
1 MHz to 25 MHz ......................................................±(10 % + 20 counts)
PWM Voltage (V Pwm)
Purpose ....................................................................To measure on pulse width modulated signals, like Motor Drive inverter outputs
Principle ....................................................................Readings show the effective voltage based on the average value of samples over a whole number of
periods of the fundamental frequency
Accuracy ................................................................... As V AC+DC for sinewave signals
Peak Voltage (V Peak)
Modes .......................................................................Max peak, Min peak, or pk-to-pk
Maximum voltage with 10:1 or 100:1 probe..............1000 V