EasyManua.ls Logo

Fluke PM6681 - Page 75

Fluke PM6681
153 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remove the probe.
-
Read
the counters
display.
-
Adjust
C40=1X
A until the displayed value has increased
5 to
1
0
mV.
10X
Attenuator
channel A
NOTE: Three digits on the display,
-
Select 10X on input A.
-
Connect the
probe to
test point X2. For trimmers and
testpoints see Figure
6-6.
-
Adjust
C1=10X A and C432
until
best
possible
pulse,
without any overshoots
or undershoots.
Adjust C1=10X A
until
the tilt of the top line Is 0.3 V.
See
Figure
6-7.
Remove the probe.
-
Read the counters display.
-
Adjust
C40=10X
A
until the displayed value has Increa-
sed
10
to
30 mV.
-
Shift between
IX
and
10X and check that the displayed
value does not differ
more than ±50 mV.
1X
Attenuator channel
B
NOTE:
Four digits on the display,
-
Connect the
probe to test point X4. For
trimmers
and
testpoints
see Figure
6-6.
Adjust C38=1X B until
the tilt
of the top line Is 0.3
V.
See Figure
6-7.
-
Remove the probe.
-
Read
the counters
display.
Adjust C38=1X B until the
displayed value
has increased
5
to
1 0
mV.
10X
Attenuator
channel B
NOTE: Three digits on the
display.
-
Select 10X on input B.
-
Connect the probe
to test
point X4. For trimmers and
testpoints see Figure
6-6.
Adjust C9=10X B and C436
until best possible pulse,
without any overshoots or
undershoots.
-
Adjust
C9=10X
B
until the tilt of the top line is 0.3 V.
See Figure
6-7.
-
Remove the
probe.
-
Read the counters
display.
-
Adjust
C38=10X
B
until the displayed value has increa-
sed 10 to 30 mV.
-
Shift between
IX
and
10X and check that the displayed
value does not differ
more than ±50 mV.
Sensitivity
Setup
PM
6681
Function DUTY FA
Input A+B
50 ti/AC/Manual trigger levels
Attenuation IX
Tripper levels 0 V
Measuring time 100 ms
LF synthesizer
Frequency 1 kHz Sinus or triangle
Amplitud 500
mV
pp
before the
attenuator
Oscilloscope
Amplitude 50
mV/div
(10:1
probe)
Time
200
jis/div
Table
6-7
Sensitivity
setup.
-
Connect
the LF synthesizer
via
an
20dB attenuator to
input A.
-
Check with the
oscilloscope
that the
signal at the
input
of
the counter is clean and real sinus or triangle.
-
Adjust the resistor R18=OFFSET A
until the
counter
shows 0.500 ±0.001.
-
Connect
the LF synthesizer via the 20dB attenuator
to
input
B.
-
Press SWAP.
-
Adjust the resistor R44=OFFSET B until the counter
shows 0.500 ±0.001.
Trigger
Levels
Setup
PM 6681
Function FREQ
A
Input A+B
1 MCi/DC/Manual
trigger
levels
Attenuation
1X
Trigger levels 5 V
Measuring time 80 ns
DC
source
Amplitude
5.0 V ±50 mV.
Table
6-8
Trigger Levels setup.
Adjustment
Channel A
Connect
a
stable DC voltage to the
A
input of
the coun-
ter via a T-piece. See Figure
6-8.
-
Connect a DMM to the other end of the T-piece.
NOTE: Use coaxial cables to avoid signal interferance.
Input Amplifier
6-7