QA-ES III
Getting Started Manual
18
Power
Ranges ............................ 0.0 W to 99.9 W
100 W to 500 W
Accuracy ...................... < 10 W: ±5 % + 1 W
10 W: ±5 %
Maximum
At 25% duty cycle (10 seconds on,
30 seconds off) ........... 10 : 300 W,
20 to 2900 : 400 W,
3000 to 5200 : 200 W
At 10% duty cycle (5 seconds on,
45 seconds off) ........... 10 : 300 W,
20 to 2400 : 500 W,
2425 to 2900 : 400 W,
3000 to 5200 : 200 W
Current
RMS ........................... 0 mA to 5,500 mA
Accuracy ..................... ±(2.5 % of reading + 1 mA)
Voltage
Peak .......................... 10 kV Peak to Peak
Accuracy ..................... ±(10 % of reading + 50 V)
Crest Factor ........................ 1.4 to 16.0
Defined as the ratio of Peak voltage to RMS voltage (Vpk /
Vrms), using the larger of the 2 peaks (positive or
negative)
Vessel Sealing Measurement
Loop Current, RMS ............. 0 mA to 5500 mA
Accuracy ............................ ±(2.5 % of reading + 1 mA)
HF Leakage Current
Fixed Load .......................... 200
Load Accuracy .................... ±2.5 %
Power rating ........................ 400 W
Additional Fixed Load ......... 200
Current, RMS ...................... 0 mA to 5500 mA
Accuracy ............................. ±(2.5 % of reading + 1 mA)
CQM Test (Contact Quality Monitor)
Resistances ........................ 0 to 475 (by 1 )
Accuracy ............................. 0 to 10 ±0.5 ,
11 and above ±5 %
Power rating ....................... 0.5 W
Auto Time interval ............... 1 second to 5 seconds
Oscilloscope Output
1 V per ampere of input current, typical.
Footswitch Simulations
Cut and Coag
Communications
USB Device Port ................. Micro B connector
Full speed
Wireless Port ...................... 802.15, Speed:115,200 baud
Memory
Test Records ...................... 5,000
Non-volatile ........................ retained through power cycling
Calibration
Traceable to the International System of Units (SI) through the
appropriate National Metrology Institutes such as NIST or
through intrinsic standards.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke QA-ES III and is the answer not in the manual?

Fluke QA-ES III Specifications

General IconGeneral
BrandFluke
ModelQA-ES III
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals