Electrosurgery Analyzer
Technical Specifications
17
Electromagnetic Compatibility (EMC)
IEC 61326-1: Basic
Emissions
Classification ...................... IEC CISPR11: Group 1, Class A.
Group 1 have intentionally generated and/or use
conductively coupled radio-frequency energy which is
necessary for the internal functioning of the equipment
itself. Class A equipment is suitable for use in non-
domestic locations and/or directly connected to a low-
voltage power supply network
USA (FCC) ......................... Intentional Radiators
This device complies with part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1)
This device may not cause harmful interference, and (2)
this device must accept any interference received,
including interference that may cause undesired operation.
(15.19)
Changes or modifications not expressly approved by Fluke
could void the user's authority to operate the equipment.
(15.21)
Safety
IEC 61010-1:Overvoltage category II, pollution degree 2
IEC 61010-2-030: Measurement 5,000 V
Wireless Module Listing
FCC (United States) compliant
(Class A)............................. FCC ID: X3ZBTMOD3
IC (Industry Canada)
compliant ............................ IC: 8828A-MOD3
CE (European) certified ...... CE0051
Technical Specifications
Specifications apply for a period of one year from date of the
most recent calibration.
Measures ................................ Cut and coag waveforms
Monopolar and bipolar outputs
Power and current
measurements ....................... true RMS
Bandwidth ............................. 30 Hz to 5 MHz at -3 dB including
loads
Delay Time for single
measurements ....................... 0.2 seconds to 4.0 seconds from
Foot Switch activation to start of
measurement
Duty Cycle
Variable Load ..................... 10 seconds on, 30 seconds off, at
100 W, all loads
Fixed 200 Load .............. 10 seconds on, 30 seconds off, at
400 W
Generator Output Measurements
Load Resistance
Variable ......................... 0 , 10 , 20 , 25 to 2500
(by 25 ), 2500 to 5200 (by
100 )
Accuracy ........................ ±2.5 %