EasyManua.ls Logo

Fluke ScopeMeter 123B - Waveform Filtering; Noise Reduction Filter; Glitch Display Options

Fluke ScopeMeter 123B
192 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
123B/124B/125B
Users Manual
Noise Reduction
To view the waveform without higher frequency noise:
1. Push to open the Scope and Meter button bar.
2. Push 3 to open the INPUT SETTINGS menu.
3. Use YZ to highlight the Noise filter as ON.
4. Push to make the change.
This noise filter is a 10 kHz bandwidth limit filter.
hxv15.eps
Glitch Display
By default, the displayed waveform shows glitches. For
each time position the minimum and maximum values
since the last time position are shown. A glitch of 25 ns or
wider is still visible on the screen even when using a
slower timebase.
To disable this feature and see one sample/channel for
each sample time:
1. Push to open the Scope and Meter button bar.
2. Push 2 to open the SCOPE menu.
3. Use
YZ to highlight the Type as Glitch Off.
This feature will be turned off for both channel A and
channel B.
4. Push to make the change.
5. Push 4 to exit the menu.
24

Table of Contents

Related product manuals