123B/124B/125B
Users Manual
Table 12. Bus Test Screen Indicators
Indicator Description
Bus activity indicators
Bus activity indicator 1:
(filled) voltage measured
(open) no voltage measured
Bus activity indicators 2 and 3:
(both open) no activity
(blinking)
activity
Busy, the Test Tool is measuring/processing data.
c No reading available.
f Test OK. Measurement results are within 80% of allowable range, see Figure 12.
d Warning. Measurement results are between 80% and 100% of allowable range, see Figure 12.
e Test failed. Measurement results are out of allowable range, see Figure 12.