Specifications
Input A
Ohm ()
Ranges
125B ................................................................ 50 , 500 , 5 k, 50 k, 500 k, 5 M, 30 M
124B, 123B ..................................................... 500 , 5 k, 50 k, 500 k, 5 M, 30 M
Accuracy..............................................................±(0.6 % + 5 counts)
50 ±(2 % + 20 counts)
Full Scale Reading:
50 to 5 M ................................................... 5000 counts
30 M ............................................................. 3000 counts
Measurement Current ......................................... 0.5 mA to 50 nA, decreases with increasing ranges
Open Circuit Voltage ........................................... <4 V
Continuity (CONT)
Beep .................................................................... <(30 ±5 ) in 50 range
Measurement Current ......................................... 0.5 mA
Detection of shorts .............................................. 1 ms
Diode
Measurement Voltage
@0.5 mA ......................................................... >2.8 V
@open circuit .................................................. <4 V
Accuracy..............................................................±(2 % + 5 counts)
Measurement Current ......................................... 0.5 mA
Polarity ................................................................ + on input A, - on COM
Capacitance (CAP)
Ranges................................................................ 50 nF, 500 nF, 5 µF, 50 µF, 500 µF
Accuracy..............................................................±(2 % + 10 counts)
77